The 753 and 754, 743 and 744 work with the Fluke software, and with popular programs from Honeywell Meridium, Emerson, Cornerstone, Yokogawa, Prime Technologies, Intergraph and others. It allows you to create procedures, instructions, and action lists to deliver fast, easy documentation.
Work smarter. Work faster.The 754 is a power multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. The 754 also features a powerful built in HART® interface capable of performing nearly all the day-to-day tasks you now perform with a separate communicator.
HART Capabilities
Versatile HART Protocol support
1Sensor Trim not supported The 754 supports the commands contained in HART protocol Version 5.7. With 2 MB of memory, the 754 supports a substantial set of HART instructions:
Pressure Modules Covers virtually any pressure application including gauge, differential, dual (compound), absolute, and vacuum.
A family of 29 optional pressure modules provides pressure calibration and measurement capabilities. Twenty-eight modules are available, with basic accuracy specs to 0.05%. Ranges start at 0 — 1 in H20 (0 — .25kPa) and go to 0 — 10,000 psi (0 — 70,000 kPa). Additional information about pressure modules is available on the Pressure Modules Home Page. Automated Procedures Allow you to quickly set up powerful, automated calibration procedures for linear transmitters, DP flow transmitters, and one- and two-point limit switches. Simply select the appropriate measure and/or source functions and fill out the procedure template. The 750 Series does the rest. It quickly performs the test, calculates the errors, and displays the final results, highlighting out of tolerance points. Custom Units Enable you to map one unit to another, such as mV to ºC or ºF. Allows you to use the Fluke 750 Series with millivolt output accessories such as the Fluke 80T-IR Temperature Probe, and to document tests using non-supported units such as parts per million or revolutions per minute. User-entered Values Enables technicians to record calibration results that were sourced and/or measured by other devices such as panel meters or readout-only devices. Limit Switch Calibration Procedures perform fast, automated calibration of one and two-point limit switches for voltage, current, temperature, and pressure. Differential Pressure Flow Instrument Calibration Routines use a square root function to directly calibrate DP flow instruments. Additional Features Calibrate temperature, pressure, voltage, current, resistance, and frequency. Since it both measures and sources, you can troubleshoot and calibrate all with one rugged tool. Powerful, yet easy to use The easy-to-follow menu-driven display guides you through any task. Get up to speed in minutes, not days. Programmable calibration routines enable you to create and run automated as-found/as-left procedures to ensure fast, consistent, calibrations. Records and documents results To support your ISO-9000 or regulatory standards the Fluke 754 captures your calibration results, eliminating the need to juggle a pen and pad in the field. The USB interface lets you transfer the results to a PC, thus saving the time of having to manually transcribe them when you return to the shop. Truly hand-held Small enough to fit easily into a tool bag and to use in tight spaces. Runs an entire shift on a rechargeable Li-ion battery pack. Rugged and reliable Count on Fluke's rugged design to deliver top accuracy and reliability in harsh environments. Overmolded urethane case stands up to rough handling in industrial environments. Bright white display Lets you read your results in any kind of light. Backlight has three (3) settings. Soft keys Provide one-touch access to enhanced functions such as task lists, automated procedures, scaling, min/max, stepping and ramping, and review memory. Three operating modes Measure, Source, or simultaneous Measure/Source, - enable technicians to troubleshoot, calibrate, or maintain instrumentation with just one tool. Multi-lingual interface Displays instructions in English, French, German, Spanish, and Italian. Built-in algebraic calculator With four functions-plus square root-stores, recalls, and performs calculations required for setting up instruments or evaluating data in the field. Use it to set the source function to a calculated value. There's no need to carry a pencil and paper or a separate calculator. Programmable measurement delay Inside automated procedures permits calibrating instruments that respond slowly. |
Measurement Accuracy | |||||||||||||||||
Voltage DC |
| ||||||||||||||||
Voltage AC |
| ||||||||||||||||
Current DC |
| ||||||||||||||||
Resistance |
| ||||||||||||||||
Frequency |
|
Source Accuracy | ||||||||||||||||
Voltage DC |
| |||||||||||||||
Current DC |
| |||||||||||||||
Resistance |
| |||||||||||||||
Frequency |
|
Technical Data | |||||||||
Data Log Functions |
| ||||||||
Ramp Functions |
| ||||||||
Loop Power Function |
| ||||||||
Step Functions |
|
Environmental Specifications | |||
Operating Temperature |
| ||
Storage Temperature |
| ||
Dust/water resistance |
| ||
Operating Altitude |
|
Safety Specifications | |||
Agency Approvals |
|
Mechanical & General Specifications | |||||||||||
Size |
| ||||||||||
Weight |
| ||||||||||
Batteries |
| ||||||||||
Battery Life |
| ||||||||||
Battery Replacement |
| ||||||||||
Side port connections |
| ||||||||||
Data storage capacity |
| ||||||||||
90 day specifications |
|
Temperature, Resistance Temperature Detectors | |||||||
Degrees or % of reading | |||||||
Type (α) | Range °C | Measure °C1 | Source current | Source °C | Allowable current2 | ||
1 year | 2 year | 1 year | 2 year | ||||
100 Ω Pt (385) | -200 to 100 100 to 800 | 0.07 °C 0.02 % + 0.05 °C | 0.14 °C 0.04 % + 0.10 °C | 1 mA | 0.05 °C 0.0125 % + 0.04 °C | 0.10 °C 0.025 % + 0.08 °C | 0.1 mA to 10 mA |
200 Ω Pt (385) | -200 to 100 100 to 630 | 0.07 °C 0.02 % + 0.05 °C | 0.14 °C 0.04 % + 0.10 °C | 500 μA | 0.06 °C 0.017 % + 0.05 °C | 0.12 °C 0.034 % + 0.10 °C | 0.1 mA to 1 mA |
500 Ω Pt (385) | -200 to 100 100 to 630 | 0.07 °C 0.02 % + 0.05 °C | 0.14 °C 0.04 % + 0.10 °C | 250 μA | 0.06 °C 0.017% + 0.05 °C | 0.12 °C 0.034 % + 0.10 °C | 0.1 mA to 1 mA |
1000 Ω Pt (385) | -200 to 100 100 to 630 | 0.07 °C 0.02 % + 0.05 °C | 0.14 °C 0.04% + 0.10 °C | 150 μA | 0.06 C 0.017 % + 0.05 °C | 0.12 C 0.034 % + 0.10 °C | 0.1 mA to 1 mA |
100 Ω Pt (3916) | -200 to 100 100 to 630 | 0.07 °C 0.02 % +0.05 °C | 0.14 °C 0.04 % +0.10 °C | 1 mA | 0.05 °C 0.0125 % + 0.04 °C | 0.10 °C 0.025 % + 0.08 °C | 0.1 mA to 10 mA |
100 Ω Pt (3926) | -200 to 100 100 to 630 | 0.08 °C 0.02 % +0.06 °C | 0.16 °C 0.04 % +0.12 °C | 1 mA | 0.05 °C 0.0125 % + 0.04 °C | 0.10 °C 0.025 % + 0.08 °C | 0.1 mA to 10 mA |
10 Ω Cu (427) | -100 to 260 | 0.2 °C | 0.4 °C | 3 mA | 0.2 °C | 0.4 °C | 0.1 mA to 10 mA |
120 Ω Ni (672) | -80 to 260 | 0.1 °C | 0.2 °C | 1 mA | 0.04 °C | 0.08 °C | 0.1 mA to 10 mA |
1For two and three-wire RTD measurements, add 0.4°C to the specifications.
2Supports pulsed transmitters and PLCs with pulse times as short as 1 ms
Temperature, Thermocouples | |||||
Type | Source °C | Measure °C | Source °C | ||
1 year | 2 years | 1 year | 2 years | ||
E | -250 to -200 | 1.3 | 2.0 | 0.6 | 0.9 |
-200 to -100 | 0.5 | 0.8 | 0.3 | 0.4 | |
-100 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
600 to 1000 | 0.4 | 0.6 | 0.2 | 0.3 | |
N | -200 to -100 | 1.0 | 1.5 | 0.6 | 0.9 |
-100 to 900 | 0.5 | 0.8 | 0.5 | 0.8 | |
900 to 1300 | 0.6 | 0.9 | 0.3 | 0.4 | |
J | -210 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 1200 | 0.5 | 0.8 | 0.3 | 0.3 | |
K | -200 to -100 | 0.7 | 1.0 | 0.4 | 0.6 |
-100 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
400 to 1200 | 0.5 | 0.8 | 0.3 | 0.4 | |
1200 to 1372 | 0.7 | 1.0 | 0.3 | 0.4 | |
T | -250 to -200 | 1.7 | 2.5 | 0.9 | 1.4 |
-200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 | |
0 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
B | 600 to 800 | 1.3 | 2.0 | 1.0 | 1.5 |
800 to 1000 | 1.0 | 1.5 | 0.8 | 1.2 | |
1000 to 1820 | 0.9 | 1.3 | 0.8 | 1.2 | |
R | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 100 | 1.5 | 2.2 | 1.1 | 1.7 | |
100 to 1767 | 1.0 | 1.5 | 0.9 | 1.4 | |
S | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 200 | 1.5 | 2.1 | 1.1 | 1.7 | |
200 to 1400 | 0.9 | 1.4 | 0.9 | 1.4 | |
1400 to 1767 | 1.1 | 1.7 | 1.0 | 1.5 | |
C | 0 to 800 | 0.6 | 0.9 | 0.6 | 0.9 |
800 to 1200 | 0.8 | 1.2 | 0.7 | 1.0 | |
1200 to 1800 | 1.1 | 1.6 | 0.9 | 1.4 | |
1800 to 2316 | 2.0 | 3.0 | 1.3 | 2.0 | |
L | -200 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 900 | 0.5 | 0.8 | 0.2 | 0.3 | |
U | -200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 |
0 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
BP | 0 to 1000 | 1.0 | 1.5 | 0.4 | 0.6 |
1000 to 2000 | 1.6 | 2.4 | 0.6 | 0.9 | |
2000 to 2500 | 2.0 | 3.0 | 0.8 | 1.2 | |
XK | -200 to 300 | 0.2 | 0.3 | 0.2 | 0.5 |
300 to 800 | 0.4 | 0.6 | 0.3 | 0.6 |
Documenting Process Calibrator-HART including:
Untuk keterangan lebih lanjut, silahkan download PDF ini :